Resources · EU MDR Class I

A Class I Declaration of Conformity is short.Nine things sit behind it.

What an ordinary EU MDR Class I device needs before you can sign it and apply the CE mark.

First: are you really Class I?

1. Is your product software that provides information used for diagnostic or treatment decisions, or monitors physiological processes?
2. Is it supplied sterile, does it have a measuring function, or is it a reusable surgical instrument?
Answer both questions to see your likely route.

Indicative only. Not a classification and not legal or regulatory advice. Classification is the manufacturer's responsibility.

What sits behind your Declaration of Conformity

EU Declaration of Conformity

Art. 19 · Annex IV. You sign it.

CE marking

Art. 20. You apply it.

1Classification and intended purposeConfirm your class and your exact claims. Everything else depends on it.Annex VIII · Rule 11
  • Your class follows from your intended purpose and the Annex VIII classification rules, not from how simple the product feels.
  • Software that provides information used for diagnostic or therapeutic decisions, or that monitors physiological processes, falls under Rule 11 and is Class IIa or higher.
  • State the intended purpose and claims precisely. Labelling, evidence and marketing all have to stay consistent with them.
2Quality management systemA documented, maintained system for design, production and change.Art. 10(9) · ISO 13485
  • Manufacturers must establish, document, implement, maintain and continually improve a quality management system proportionate to the risk class and type of device.
  • It has to cover, among other things, regulatory strategy, design and development, supplier control, risk management, clinical evaluation, post-market surveillance, incident reporting and corrective and preventive action.
  • ISO 13485 is the usual way to structure it. For a self-declared Class I device a certificate is not required, but the system still has to exist and work.
3Risk managementHazards identified, controlled and kept up to date across the lifecycle.Annex I · ISO 14971
  • A documented risk management system runs through the whole lifecycle: identify hazards, estimate and evaluate risks, control them, and review the results against real-world data.
  • Risk controls have to be verified, and the benefit-risk balance has to stay acceptable.
  • For software, risk management also connects to the software lifecycle (IEC 62304 clause 7) and to cybersecurity.
4Technical documentationThe evidence file a regulator can inspect on request.Annexes II and III
  • Annex II sets out the file: device description and specification, information supplied with the device, design and manufacturing information, and the general safety and performance requirements (GSPR) checklist.
  • It also holds the benefit-risk analysis and risk management file, and the verification and validation evidence, including the clinical evaluation. For software this normally includes lifecycle evidence to IEC 62304.
  • Annex III post-market surveillance documentation sits alongside it. Keep it current and ready for a competent authority to request.
5Clinical evaluationProof the device performs and is safe for its intended purpose.Art. 61 · Annex XIV
  • Required for every class: a planned clinical evaluation, based on clinical data, showing conformity with the relevant general safety and performance requirements.
  • Data can come from the device itself, from an equivalent device where strict equivalence conditions are met, from the literature, or from clinical investigations where needed.
  • It is updated throughout the lifecycle, using post-market data.
6Labelling and instructionsWhat the user sees and is told, in the right languages.Annex I, Chapter III
  • The label and instructions for use must carry the information listed in Annex I, Section 23.
  • Language requirements are set by each Member State where you place the device on the market.
  • Claims on labels, instructions and marketing have to match your intended purpose and your clinical evidence.
7Responsible person and coverA named person responsible for compliance, plus liability cover.Art. 15 · Art. 10(16)
  • You must have a person responsible for regulatory compliance (PRRC) with the required expertise. Micro and small enterprises may instead have one permanently and continuously at their disposal.
  • The PRRC makes sure conformity is checked before release, that the technical documentation and Declaration of Conformity are drawn up and kept up to date, and that post-market and vigilance obligations are met.
  • You also need measures for financial coverage of liability for defective devices, proportionate to risk class, device type and company size (Art. 10(16)).
8Post-market surveillancePlan, collect, report and act on real-world data.Arts. 83 to 92
  • A post-market surveillance system and plan, proportionate to the risk class and appropriate for the device (Arts. 83 to 84, Annex III).
  • For Class I, a post-market surveillance report summarises the results and is updated when necessary (Art. 85). Periodic safety update reports apply to higher classes (Art. 86).
  • Serious incidents and field safety corrective actions must be reported to the authorities within the set timelines (Arts. 87 to 92).
9UDI and EUDAMEDUnique identification and registration before you place it on the market.Arts. 27 to 31
  • Assign a UDI to the device and apply the UDI carrier where required. Software has its own UDI rules.
  • Register yourself as an actor in EUDAMED (single registration number) and register the device and its UDI-DI there before you place it on the market.
  • EUDAMED is being phased in. Check the current position in the sources and currency notes below.
Based outside the EU or EEA, including the UK? You also need an EU authorised representative (Art. 11), and importers and distributors have duties of their own.

Class IIa, IIb and III are coming next.

Sources and currency

Current as of 9 November 2026. Verify before relying.

EUDAMED: Commission Decision (EU) 2025/2371, published in the Official Journal on 27 November 2025, made the first four EUDAMED modules (actor registration, UDI and device registration, notified bodies and certificates, market surveillance) mandatory from 28 May 2026. For devices already placed on the market before that date, registration is due within 12 months of the notice, which the Commission's transition chart gives as 28 November 2026.
European Commission: EUDAMED mandatory use

MDR/IVDR revision: The Commission proposed a targeted revision of the MDR and IVDR on 16 December 2025 (COM(2025) 1023). It is still being negotiated by the European Parliament and Council and is not expected to be adopted before 2027. Regulation (EU) 2017/745 as it stands applies today.

UK: UK rules (MHRA) are separate and not covered here.

Primary references

Reference model · November 2026. This page is an illustrative reference developed by MediLife Digital Health specialists. It does not replace a device-specific regulatory assessment and is not legal or regulatory advice. Verify current requirements and national rules before relying on it.

Know which of the nine are not in place yet?

A readiness review of your actual documentation shows where the gaps are and what to do first.

Discuss Your Regulatory Readiness

This page is a reference model: it does not assess your documentation or perform any automated compliance check.